STANDARD MODE TREATMENT
NOTE: While using the Controller, should an in-process iontophoresis treatment need to be stopped or paused, DO NOT suddenly
remove the Controller from the Patch without first switching off the Controller. To stop or pause a treatment while the Controller is
administering iontophoresis, press the ON/OFF button and wait a few moments for the Controller to turn off.
NOTE: For additional information on the Controller, refer to the Controller Instructions for Use.
1. Press the ON/OFF button on the Controller. The Green I-Bresis™ Light will blink slowly.
2. Push the Standard Mode button on the Controller. The 2mA indicator light will blink slowly. Each additional depression of the button
will scroll to the next setting—3mA or 4mA.
3. Attach the Controller to the Patch. The Patch connector (located at the center of the Patch) plugs into the slot on the back of
the Controller. Ensure that the Patch connector is fully and securely engaged into the Controller—a click will be heard upon full
engagement.
4. Position the patient so that there is no pressure on the Patch during treatment.
5. Press the START button to begin treatment. The Green mA Light will blink more rapidly, then glow steadily.
6. To change the iontophoresis current setting while the Controller is administering a treatment, press the Standard Mode button to
re-select the desired setting. Within a few moments, the Controller will automatically adjust to the new setting.
7. In 10–20 minutes (see following table) the Controller will sound a beep and the lights will turn off, indicating the 40mA-minute
treatment has been completed.
Standard Mode 40 mA-min 80 mA-min
2mA 20 minutes Repeat steps 1–2 and 5–7
3mA 13 minutes Repeat steps 1–2 and 5–7
4mA 10 minutes Repeat steps 1–2 and 5–7
8. Remove the Controller from the Patch.
9. Remove and discard the Patch after the treatment has been completed. The Patch cannot be reused.
8. The average time to complete the dose is indicated in the following table. To prevent excessive dosing, the Patch automatically
switches off iontophoresis after the maximum dose has been administered.
I-Bresis™ Mode 40 mA-minutes 60 mA-minutes 80 mA-minutes
Wear Time 1 hour 1.5 hours 2 hours
9. Instruct the patient to remove and discard the Patch after one to two hours for a 40 to 80 mA-minute dose respectively.
10. The Patch should be discarded in a receptacle out of the reach of children and pets. The Patch cannot be reused.