Separate the membrane of the electrode.
Insert a rubber packing up to the membrane connection of the electrode.
Make sure that the rubber packing should be in the right direction.
•
Fill 1mL of 0.5M NaOH into the membrane and assemble it into the
electrode inserted with the rubber packing.
Refer to 11-4 How to inject solution.
•
Assemble the electrode to the electrode holder. Insert a rubber packing into
the electrode holder.
Put a washer (P.E) on top of the rubber packing.
Assemble the electrode protection tube.
•
The electrode assembly is completed by inserting the vents of the electrode
holder and the rubber stopper aligned each other.
Caution
Be careful not to apply excessive force when handling the electrode. If the glass electrode is
broken by excessive force, it may cause injury.