Cause & Solution of Problem
Problem Number of Cause & Solution
A E.01 on screen (Circuit board is not connected) 1, 2
B E.02 on screen (Electrode is not calibrated) 3, 4, 9
C Reading on screen is not changed 3, 4, 5, 6, 7, 9
D Measuring is difcult because reading is not steady 3, 5, 6, 7, 9
E
The displayed temp. is far different from the actual
temperature. (over ±5℃)
8
NO Cause Solution
1 Faulty circuit board connection Repair the controller
2 Damaged circuit board Repair the controller
3 Damaged electrodes & Electrode’s aging Exchange electrode
4 Shortage of Electrode solution Supplement of internal liquid
5 Faulty connection of controller’s terminal Connecting terminal after removing obstacle
6 Fault of liquid to be measured Exchange the measuring liquid
7
The uid velocity of liquid to be measured is
not steady
Makes the uid velocity steady
8 Fault of temp. compensation circuit Exchange temp. compensation PCB
9
Damaged electrode membrane & electrode
membrane’s aging
Exchange electrode membrane