Programmable DC Power Supply 62000L Series User’s Manual
4-12
Set dwelling of a step. The step index is from S0 to S7. Maximum dwelling is 86399999ms.
OUTPut:SEQuence:STEP:DWELI? {<step>}
Query dwelling of a step. The step index is from is S0 to S7. Return dwelling in ms.
OUTPut:SEQuence:STEP? {<step>}
Query a step setting. Step is S0~S7.
4.2.4 System-Related Commands
The following system-related commands are used to display the power supply’s status. For
detailed command information, please read the command descriptions coming next to the
commands.
DISPlay[:WINDow]
[:STATe] {OFF|ON}
[:STATe]?
:TEXT[:DATA] <quoted string>
:TEXT[:DATA]?
:TEXT:CLEar
SYSTem
:BEEPer[:IMMediate]
:BEEPer:ALARm:OVP[:STATe] {0|1|OFF|ON}
:BEEPer:ALARm:OVP[:STATe]?
:BEEPer:ALARm:OCP[:STATe] {0|1|OFF|ON}
:BEEPer:ALARm:OCP[:STATe]?
:BEEPer:NORMal[:STATe] {0|1|OFF|ON}
:BEEPer:NORMal[:STATe]?
:ERRor?
:VERSion?
*IDN?
*RST
*TST?
*SAV {0|1|2|3}
*RCL {0|1|2|3}
DISPLay {OFF|ON}
This command is used to turn the front-panel display off or on. When turning off the display,
the display will show nothing, but ERR annunciator. After returning to the local mode by
pressing LOCAL key on the front panel, the display will be turned on automatically.
DISPlay?
This query command is used to query the display ON/OFF status. If 0 is returned, it means the
display is under OFF status. If 1 is returned, it means the display is under ON status.
DISPlay:TEXT <quoted string>
This command is used to set a string to display on the front panel. Once the command is
executed, the front panel display will be turned OFF, and the string will be shown on the
display. The string to set can only contain up to 49 alphanumeric characters and must in
enclosed quotes. The string will scroll on the display if it is longer than 6 characters. ALL
letters displaying on the display will replace with uppercase ones and characters that can not