EasyManua.ls Logo

Cima PLUS - Control Unit

Cima PLUS
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Plus 42 - 45 - 50 - 55
09-2008
30
ELECTRICAL
CONTROL
PANEL
TO THE
SERVICE
SOCKET
MOUNTING
BRACKET
E10
E9
®
+
-
ON
R
e
m
o
t
e
C
o
n
t
r
o
l
HO
2
ON
OFF
OFF
Pump
00085
TO THE ELECTRICAL
DISTRIBUTOR (E9)
W2
W1
00086
IN CASE OF A PIPE BURST:
1 - Stop the tractor and remove the key from the control panel.
2 - Close the electrovalves (E9), the switches on the control panel (E10) must be on OFF.
3 - Place the lever of the 3-way tap (P6) in position "d".
4 - Replace the broken pipe (for the elimination of these parts, please follow the current regulations).
5.3.3 - E.P.A. Control Unit
Act in a similar way as indicated for installing the electric control power plant (E10)
m
.
A
U
T
M
A
N
O
N
P
R
O
C
L
R
ON
O
F
F
0
L
H
A
C
O
N
F
IG
2
MOUNTING
BRACKET
E.P.A.
CONTROL UNIT
00930
(see Paragraph 5.3.2.)
Position the electric cables in such
a way to prevent them from getting
torn during the treatment.
09-2008
Plus 42 - 45 - 50 - 55
39
The filter must always be below the level of the water to be loaded. The piping must
never be above the pump’s suction point and must never be excessively bent in
the vertical direction.
8. Through the secondary filler BS, pour approximately 15~20 litres of water into the tank: the quantity
must be sufficient to FILL the piping and ACTIVATE the pump to start the recycle.
9. Activate the power take off (PTO) and take it to a rotation speed of at least 500 RPM.
10. When the pump has started the circulation of the water poured, close tap (P2) (o P15) (lever on
“c”) to start up the filling.
11a. WITH MEDICINAL MIXTURES OR LIQUID PRODUCTS (WITHOUT MIXER DEVICE).
11a1 Pour the mixture of medicinal envisaged through the main filler (BP), together with the water
used to wash the canister of product and the tools used in the preparation.
11a2. Close the main filler (BP) of the tank.
11a3. After completing the filling, disengage the PTO
11a4. Close the tap (P6) setting the lever in position “c”.
11a5. Disconnect the filler piping and replace the stopper on the pump.
11a6. Open the faucet (P2) (or (p15) by putting the lever on position “a”
11a7. Open the tap (P6) setting the lever in position “a”
11b. WITH POWDER PRODUCTS: UTILISING THE MIXTURE DEVICE
11b1. Pour the dose of medicinal in powder form foreseen for each loading into the main filler (BP)
of the tank.
11b2. Close the main filler (BP) of the tank.
11b3. Open the cock (P12) (lever on “a”).
11b4. Wait for the filling to be completed.
11b5. Close the cock (P12) (lever on “c”).
11b6. Disengage the power take off (PTO).
11b7. Close the faucet (P6) (lever on “c”)
11b8. Disconnect the filler piping and replace the stopper on the pump.
11b9. Open the faucet (P2) (or (p15) by putting the lever on position “a”
11b10.Open the tap (P6) setting the lever in position “a”
12. Engage the Power Takeoff and move to the treatment place, keeping the Power Takeoff running
rate at least 500 RPM during all the way.
IN THE AREA OF THE TREATMENT
13. Disengage the power take off (PTO) and rest the machine on the ground.
14. STOP THE TRACTOR AND REMOVE THE KEY FROM THE CONTROL PANEL.
15. Make sure that the fan is stopped. Hook the fan to the servo amplifier as indicated at point 8.1.b.
16. Start the tractor, always observing the safety norms.
17. Lift the unit. Engage the Power Takeoff and set it up to a running rate of at least 500 r.p.m., in order
to carry out a further agitation.
18. CARRY OUT THE TREATMENT (See the paragraph 10.3).

Other manuals for Cima PLUS