Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Bone Growth Stimulator is intended for use in the electromagnetic environment specified
below. The customer or the user of the Bone Growth Stimulator should assure that it is used in
such an environment.
Emissions
Tests
Compliance Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 1
The Bone Growth Stimulator uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B The Bone Growth Stimulator is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
N/A N/A
Voltage
Fluctuations
IEC 61000-3-3
N/A N/A
Declarations EMC Tables
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES for
RF Emissions Class B