8.3 Performance characteristics
General
Processing delay <6 ms
Measurements according to IEC60118-9 2019 2
nd
Ed�
(conforms to ANSI S3�22-2014 (R2020))
Output Vibratory Force Level (re. 1μN)
(90 dB SPL input, FOG)
Max 121 dB
HFA 113 dB
ASML frequency response curve Max 48 dB
HFA 42 dB
RATMSL
1
level (re.1 μN/ 20 μPa)
(60 dB SPL input, RTS)
1
HFA 35 dB
Frequency range
1
200-9850 Hz
Equivalent input noise
1
24 dB SPL
Battery current
1
(65 dB Input SPL (rel. 20μPa) at 1kHz)
1.8 mA
Total Harmonic Distortion
1
70 dB SPL at 500 Hz < 3%
70 dB SPL at 800 Hz < 0.3%
65 dB at 1600 Hz < 0.3%
65 dB at 3200 Hz < 0.3%
Measurements according to IEC60118-0 2022 4
th
Ed�
(conforms to ANSI S3�22-2014 (R2020))
AGC characteristics
1 2
Attack time 12 ms
Release time 70 ms
1
Performed with reference test setting
of the gain control (RTS) according to
IEC60118-9:2019 2nd Ed. The RTS is set
to full-on-gain minus 7 dB and expansion
active.
2
Measured at 2 kHz with maximum
compression ratio.
AMSL = Full-on acousto-mechanical
sensitivity level re. 1 μN/20 μPa
FOG = Full On Gain
RTS = Reference Test Setting
HFA = High Frequency Average
SPL = Sound Pressure Level re. 20 μPa
RTAMSL = Reference test acousto-mechanical
sensitivity level re. 1 μN/20 μPa