21
English
Output vibratory force level [dB re. 1µN]
Output vibratory force level frequency responses
Typical output vibratory force level frequency response
(Measured on skull simulator)
Figure 1: OVFL60 1 and OVFL90 frequency response curves for Baha 6 Max
1
Performed with reference test setting of the gain control (RTS) according
to IEC60118-9:2019 2
nd
Ed. The RTS is set to full-on-gain minus 7 dB and
expansion active.
AMSL [dB re. 1 μN / 20 μPa]
Acousto-mechanical sensitivity level frequency response
AMSL
Typical AMSL (Full-on acousto-mechanical sensitivity level
frequency response) (Measured on skull simulator)
Figure 2: AMSL (Full-on acousto-mechanical sensitivity level
frequency response) curve for Baha 6 Max