Alarm for unacceptable water conductivity.
Alarm for lack of water in the clean water tank.
Alarm for failure to fill chamber with required water.
Indicates steam leakage from the chamber.
Indicates improper temperature increase during steam phase.
Indicates chamber temperature exceeded 150°C.
Indicates pressure did not drop correctly during pre-vacuum.
Indicates vacuum was not created properly.
Indicates temperature is outside sterilisation tolerance.
Indicates inconsistent operation of level probes.
Indicates an open or interrupted coil element circuit.
Alarm related to the automatic water inlet system.
Indicates the door is not correctly closed or locked.
Triggered by disconnected or faulty PT1 temperature probe.
Indicates incorrect pressure drop during draining.
Indicates chamber temperature doesn't match theoretic steam temp.
Triggered by disconnected or faulty pressure transducer.
| Type | Autoclave |
|---|---|
| Chamber Volume | 18 liters |
| Sterilization Temperature | 121°C to 134°C |
| Power Supply | 220-240V, 50/60Hz |
| Power Consumption | 2.5 kW |
| Material | Stainless Steel |
| Safety Features | door lock |
| Display | LCD |
| Data Logging | USB port for data export |