Minimum recommended sterilization exposure times are as follows:
Table 1: Sterilization Parameters
Sterilization Type Temperature
Minimum
Exposure
Time
Dry Cycle
Sterilization Guidelines for Sterilizing PowerPro Attachments Individually or in the
PowerPro PRO5090 System Sterilization Case
Steam Pre-vacuum 270 - 272°F (132 - 133°C) 4 minutes 8 minutes minimum *
Steam Gravity 270 - 272°F (132 - 133°C) 10 minutes 8 minutes minimum *
Steam Gravity 250 - 254°F (121 - 123°C) 40 minutes 8 minutes minimum *
Sterilization Guidelines for Sterilizing P
owerPro Attachments in the PowerPro Basic
Sterilization Case (Lid - PRO2091/Tray - PRO2092)
Steam Pre-vacuum 270 - 272°F (132 - 133°C) 4 minutes 8 minutes minimum *
Steam Gravity 270 - 272°F (132 - 133°C) 15 minutes 8 minutes minimum *
Steam Gravity 250 - 254°F (121 - 123°C) 40 minutes 8 minutes minimum *
Sterilization Guidelines for Sterilizing PowerPro Attachments in the PowerPro Complete
Sterilization Case (PRO2090)
Steam Pre-vacuum 270 - 272°F (132 - 133°C) 4 minutes 10 minutes minimum *
Steam Gravity 270 - 272°F (132 - 133°C) 15 minutes 10 minutes minimum *
Steam Gravity 250 - 254°F (121 - 123°C) 40 minutes 10 minutes minimum *
* CAUTION: The recommended dry cycle must be run on all attachments every time they
are sterilized. Failure to use a dry cycle may lead to reduced product performance or prema-
ture product failure. Operation of an attachment that is not completely cool and dry may
decrease performance and/or reliability.