5.1 Main unit
Cleaning may be required as a result of spilled drops of blood or when required after
inspection.
If spillage occurs, the unit must immediately be removed from service and cleaned com-
pletely before resuming use.
Use a soft lint-free tissue, moistened with a mild detergent to clean the outside of the
Main unit.
5.2 Hand Unit
Cleaning may be required as a result of spilled drops of blood or when required after
inspection.
For cleaning the electrodes you should remove the cover rst and then remove the trig-
ger by rstly pushing the front and then lift the trigger to remove. Clean the tube and
both electrodes with a soft lintfree tissue moistened with mild detergent. Dry carefully
and ensure that the electrodes are completely dry to prevent sparks.
Note! After cleaning inspect the electrode for damage.
Assemble the parts in reverse order.
Note! We recommend some sealing tests before resuming use, ensure that the
electrode moves smoothly in the hand unit.
Caution! Do not dip the units in liquid, as they are not waterproof. Intruding liquid will
cause malfunction, tiny arcs and reduce the lifetime use.
24
CS624 Qseal-extra Operator´s manual