Chapter21 NIBP Specification
Name Electronic Sphygmomanometer
Display 2.8'' color LCD Display
NIBP Specifications
Measurement Method Oscillometric method
Working modes Automatic
Measurement Range
Pressure
adult
0~290mmHg(0~38.6kPa)
pediatric
0~235mmHg(0~31.3kPa)
neonatal
0~140mmHg(0~18.6kPa)
Pulse: 40~240/min
Inflation
adult 160mmHg
pediatric 120mmHg
neonatal 70mmHg
Alarm Range
adult mode
SYS ALM: 40~270 mmHg
DIA ALM: 10~215 mmHg
pediatric mode
SYS ALM: 40~200 mmHg
DIA ALM: 10~150 mmHg
neonatal mode
SYS ALM: 40~135 mmHg
DIA ALM: 10~100 mmHg
Overpressure protect
adult mode 295±5mmHg
pediatric mode
240±5mmHg
neonatal mode
145±5mmHg
Resolution
Pressure 1mmHg
Measurement Accuracy
Cuff Pressure Accuracy ±3mmHg
Error
The BP Value of the device is equivalence with the measurement
value of Stethoscope. The error meets all the conditions in the
ANSI/AAMI SP-10:2002+A1:2003 +A2:2006.
Operating
Temperature/ Humidity
+5ºC~40 ºC . 15%RH~80%RH
Transport and Storage
Temperature/Humidity
-20ºC~+55ºC . ≤95%RH
Atmospheric pressure 80kPa~105kPa
Battery 4 "AA" alkaline batteries, AC Adapter separately sold
Main Unit Dimensions 130(L)*110(W)*80mm(H)
Main Unit Weight 300gram