Figure 7 Put finger in position
7 Operating Guide
1) Insert the two batteries properly to the direction, and then replace the cover.
2) Open the clip as shown in Figure 7.
3) Let the user’s finger put into the rubber cushions of the clip (make sure the
finger is in the right position), and then clip the finger.
4) Press the switch button once on front panel.
5) Do not shake the finger and keep the user at ease during the process.
Meanwhile, human body is not recommended in movement status.
6) Get the information directly from screen display.
7) The button has two functions.When the device is in standby mode, pressing
the button can exit it; When the device is in operation status, pressing the button
long can change brightness of the screen.
8) The device could change display direction according to the handing
direction.
Fingernails and the luminescent tube should be on the same side.
8 Repairing and Maintenance
Please change the batteries when the low-voltage displayed on the screen.
Please clean the surface of the device before using. Wipe the device with
medical alcohol first, and then let it dry in air or clean it by dry clean
fabric.
Using the medical alcohol to disinfect the product after use, prevent from
cross infection for next time use.
Please take out the batteries if the oximeter is not in use for a long time.
The packed device can be transported by ordinary conveyance or according
to transport contract.The device can not be transported mixed with toxic,
harmful, corrosive material.
The best storage environment of the device is - 40ºC to 60ºC ambient
temperature and not higher than 95% relative humidity.
Users are advised to calibrate the device termly (or according to the
calibrating program of hospital). It also can be performed at the
state-appointed agent or just contact us for calibration.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid.
It is recommended that the device should be kept in a dry
environment. Humidity may reduce the useful life of the device, or
even damage it.
9 Troubleshooting
The SpO
2
and
Pulse Rate
can not be
displayed
normally
1. The finger is not properly
positioned.
2. The user’s SpO
2
is too
low to be detected.
1. Place the finger
properly and try again.
2. Try again; Go to a
hospital for a diagnosis
if you are sure the
device works all right.
The SpO
2
and
Pulse Rate are
not displayed
stably
1. The finger is not placed
inside deep enough.
2. The finger is shaking or
the user is moving.
1. Place the finger
properly and try again.
2. Let the user keep
calm
The device
can not be
turned on
1. The batteries are drained
or almost drained.
2. The batteries are not
inserted properly.
3. The malfunction of the
device.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the
local service center.
The display is
off suddenly
1. The product will enter
standby mode when no
signal is in the product
within 5 seconds
2. The batteries are almost
drained.
1. Normal.
2. Change batteries.
10 Key of Symbols
Refer to instruction manual/booklet
The pulse oxygen saturation(%)
The battery voltage indication is deficient (change the
battery in time avoiding the inexact measure)
1. No finger inserted
2. An indicator of signal inadequacy
Battery positive electrode
Battery negative electrode;
1.Exit standby mode.
2.Change brightness of the screen.
Storage and Transport Temperature limitation
Storage and Transport Humidity limitation
Storage and Transport Atmospheric pressure limitation
Fragile, handle with care
11 Function Specification
The Pulse Oxygen Saturation
(SpO
2
)
Pulse Intensity (bar-graph)
SpO
2
Parameter Specification
0% ~ 100%, (the resolution is 1%).
70% ~ 100%:±2%, Below 70% unspecified.
Red light (wavelength is 660 nm)
Infrared (wavelength is 880 nm)
Pulse Parameter Specification
30 bpm ~ 250 bpm (the resolution is 1 bpm)
± 2 bpm or±2% select larger
Continuous bar-graph display, the higher
display indicate the stronger pulse.
1.5V (AAA size) alkaline batteries × 2 or rechargeable battery
Two batteries can work continually for 20 hours
About 50 g (with the batteries)
Appendix
Guidance and manufacture’s declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration –electromagnetic emission
The CMS50DA Pulse Oximeter is tended for use in the electromagnetic
environment specified below. The customer of the user of the CMS50DA Pulse
Oximeter should assure that it isused in such an environment.
Electromagnetic
environment-guidance
The CMS50DA Pulse Oximeter uses RF
energy only for their internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
The CMS50DA Pulse Oximeter is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emission
IEC 61000-3-3
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The CMS50DA Pulse Oximeter is intended for use in the electromagnetic
environment specified specified below. The the user of CMS50DA Pulse
Oximeter should assure that it is used in such an environment.
Electromagnetic
environment-guidance
Electrostat
ic
discharge
(ESD) IEC
61000-4-2
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Power
frequency
(50Hz)
magnetic
field
IEC
61000-4-8
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment
Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The CMS50DA Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of CMS50DA Pulse
Oximeter should assure that it is used in such an environment.
Electromagnetic environment
-guidance
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the CMS50DA
Pulse Oximeter
The CMS50DA Pulse Oximeter is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The customer or
the user of the CMS50DA Pulse Oximeter can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the CMS50DA Pulse
Oximeter as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of
transmitter
(m)
For transmitters rated at a maximum output power not listed above, the
recommended separation distanced in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Radiat
ed RF
ICE
61000-
4-3
Portable and mobile RF communication
equipment should be used no closer to any
part of the CMS50DA Pulse Oximeter,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
recommended separation distance
800MHz to 2.5GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcastcannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which The CMS50DA Pulse
Oximeter is used exceeds the applicable RF compliance level above, the
CMS50DA Pulse Oximeter should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the CMS50DA Pulse Oximeter.
b. Over the frequency range 150 KHz to 80 MHz, field strengths should be
less than 3V/m.