34
CONTOUR
®
TS meter CONTOUR
®
TS test strips
Technical,
Service & Care
Technical Information
Technical Information
The
CONTOUR
®
TS
blood glucose monitoring system
was tested in 100 capillary blood samples using 600
CONTOUR
®
TS
test strips. Two replicates were tested with
each of 3 lots of
CONTOUR TS
test strips for a total of 600
readings. Results were compared to the YSI
®
glucose analyzer,
which is traceable to the CDC hexokinase method. The tables
below compare the performance of the 2 methods.
Table 1 – System accuracy results for glucose
concentration < 5.55 mmol/L
Difference range
in values between
YSI laboratory
reference method and
CONTOUR TS meter
Within ±
0.28 mmol/L
Within ±
0.56 mmol/L
Within ±
0.83 mmol/L
Number (and percent)
of samples within
specifi ed range
77 of 180
(42.8%)
145 of 180
(80.6%)
170 of 180
(94.4%)
Table 2 – System accuracy results for glucose
concentration 5.55 mmol/L
Difference range in values
between YSI laboratory
reference method and
CONTOUR TS meter
Within ±
5%
Within ±
10%
Within ±
15%
Number (and percent) of
samples within specifi ed
range
297 of 420
(70.7%)
399 of 420
(95.0%)
416 of 420
(99.0%)
Accuracy
5718657_CntrTS_Karajishi_UG_EN_FpBp_v0.indd 34 3/24/17 4:13 AM
35
Technical,
Service & Care
Technical Information
Table 3 – System accuracy results for glucose
concentrations between 2.0 and 24.6 mmol/L
Within ±0.83 mmol/L or ±15% 586 of 600 (97.7%)
Acceptance criteria in ISO 15197: 2013 are that
95% of the measured glucose values shall fall within
either ±0.83mmol/L of the average measured values
of the reference measurement procedure at glucose
concentrations <5.55 mmol/L or within ±15% at
glucose concentrations 5.55mmol/L.
User Accuracy
A study evaluating glucose values from fingertip capillary
blood samples obtained by 132 lay persons showed
the following results: 100% within ±0.83mmol/L of the
medical laboratory values at glucose concentrations below
5.55mmol/L and 95.6% within ±15% of the medical
laboratory glucose concentrations at or above 5.55 mmol/L.
Precision
A repeatability study was conducted with the
CONTOUR
®
TS
blood glucose monitoring system using 5 venous
whole blood specimens with glucose levels from 2.2 to
18.3mmol/L. Multiple replicates (n=300) were tested using
multiple
CONTOUR TS
blood glucose meters and 3lots
of
CONTOUR
®
TS
blood glucose test strips. The following
precision results were obtained.
5718657_CntrTS_Karajishi_UG_EN_FpBp_v1.indd 35 4/3/17 9:17 AM
35
Technical,
Service & Care
Technical Information
Table 3 – System accuracy results for glucose
concentrations between 2.0 and 24.6 mmol/L
Within ±0.83 mmol/L or ±15% 586 of 600 (97.7%)
Acceptance criteria in ISO 15197: 2013 are that
95% of the measured glucose values shall fall within
either ±0.83mmol/L of the average measured values
of the reference measurement procedure at glucose
concentrations <5.55 mmol/L or within ±15% at
glucose concentrations 5.55mmol/L.
User Accuracy
A study evaluating glucose values from fingertip capillary
blood samples obtained by 132 lay persons showed
the following results: 100% within ±0.83mmol/L of the
medical laboratory values at glucose concentrations below
5.55mmol/L and 95.6% within ±15% of the medical
laboratory glucose concentrations at or above 5.55 mmol/L.
Precision
A repeatability study was conducted with the
CONTOUR
®
TS
blood glucose monitoring system using 5 venous
whole blood specimens with glucose levels from 2.2 to
18.3mmol/L. Multiple replicates (n=300) were tested using
multiple
CONTOUR TS
blood glucose meters and 3lots
of
CONTOUR
®
TS
blood glucose test strips. The following
precision results were obtained.
5718657_CntrTS_Karajishi_UG_EN_FpBp_v1.indd 35 4/3/17 9:17 AM
34
CONTOUR
®
TS meter CONTOUR
®
TS test strips
Technical,
Service & Care
Technical Information
Technical Information
The
CONTOUR
®
TS
blood glucose monitoring system
was tested in 100 capillary blood samples using 600
CONTOUR
®
TS
test strips. Two replicates were tested with
each of 3 lots of
CONTOUR TS
test strips for a total of 600
readings. Results were compared to the YSI
®
glucose analyzer,
which is traceable to the CDC hexokinase method. The tables
below compare the performance of the 2 methods.
Table 1 – System accuracy results for glucose
concentration < 5.55 mmol/L
Difference range
in values between
YSI laboratory
reference method and
CONTOUR TS meter
Within ±
0.28 mmol/L
Within ±
0.56 mmol/L
Within ±
0.83 mmol/L
Number (and percent)
of samples within
specifi ed range
77 of 180
(42.8%)
145 of 180
(80.6%)
170 of 180
(94.4%)
Table 2 – System accuracy results for glucose
concentration 5.55 mmol/L
Difference range in values
between YSI laboratory
reference method and
CONTOUR TS meter
Within ±
5%
Within ±
10%
Within ±
15%
Number (and percent) of
samples within specifi ed
range
297 of 420
(70.7%)
399 of 420
(95.0%)
416 of 420
(99.0%)
Accuracy
5718657_CntrTS_Karajishi_UG_EN_FpBp_v0.indd 34 3/24/17 4:13 AM
85718657_CntrTS_Karajishi_UG_EN_FpBp_v4_placed.pdf:36

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contour TS and is the answer not in the manual?

Contour TS Specifications

General IconGeneral
Sample Size0.6 μL
Test Time5 seconds
CodingNo coding required
Memory250 test results
Battery LifeApproximately 1000 tests
Measuring Range0.6 to 33.3 mmol/L
Battery1 CR2032 battery
Operating Temperature5°C - 45°C (41°F - 113°F)
Weight56.7 g (2 oz)