202D-M-V3.1
Programming
17
Step Comment
CAL 7
Decimal Point for Total Display.
The totals can be displayed with 0, 1, 2 or 3 decimal points.
CAL 8
Total Conversion Factor - whole numbers.
CAL 9
Total Conversion Factor - digits after the decimal point.
The total conversion factor enables the rate to be displayed
in one engineering unit and the totals to be displayed in
another engineering unit. The total conversion factor can
be programmed in the range of 0.000001 - 999,999.
Set to 1.000 if totals and rate are in the same unit, e.g. litres.
See Section 3.5.
CAL 10
Low Alarm - whole numbers.
CAL 11
Low Alarm - digits after the decimal point.
CAL 10 to CAL 13 are displayed regardless of whether the
high and low alarms are installed. If the high and low
alarms are not installed, these parameters can be ignored.
CAL 10 & 11 program the flow rate below which the
low alarm relay will close. The value can be
programmed in the range 0.000000 to 999,999.
CAL 12
High Alarm or Pulse Output Factor - whole numbers.
CAL 13
High Alarm or Pulse Output Factor - digits after the decimal
point.
CAL 12 & 13 program the flow rate above which the high
alarm relay will close. The value can be programmed in
the range 0.000000 to 999,999.
If the scaled pulse output is selected (see CAL0 = 1), then
the value will represent the total per pulse, e.g. 5 litres
per pulse.