12
d. Click the Dosing Time column of Step 2, input 10, then click Enter to
confirm inputted data.
When the program entering the holding stage of step 2, at after 10 minutes
the machine will inform operator to dose the chemical into beaker by
buzzer and message display.
Step 3: Target Temperature =40, Gradient = 4/min, Holding Time = 0min
a. Click the Target Temp. column of Step 3, input 40, then click Enter to
confirm inputted data.
b. Click the Gradient column of Step 3, input 4.0, then click Enter to confirm
inputted data.
If you input 999 at the holding time field, it means that machine will enter
into the Constant dyeing mode and the temperature will be kept as the
target temperature you specified.
If you input 999 at the dosing time field, it means while machine is entering
into the holding time stage and dosing chemical alarm will be activated at
once.