TROUBLESHOOTING
PROBLEM
Monitor
connected
to
AC
power,
green
Power
On
light
does
not
illuminate
upon
power
switch
depression.
“HLP”
appears
on
front
panel
display.
Recorder
does
not
function.
(Record
light
off)
(Record
light
on)
ULTRASOUND
not
functioning
properly.
Tocotransducer
not
record-
ing
contractions.
PROBABLE
CAUSE"
Monitor
not
connected
to
AC-line
receptacle.
AC
power
cord
defective.
AC
outlet
defective.
Power
cord
not
connected
to
Model
145.
Monitor
fails
self
test.
Recorder
not
activated.
Recorder
out
of
paper.
Connector
not
properly
connected
to
monitor.
Transducer
misplaced.
Too
little
gel
applied
to
transducer.
Defective
transducer.
Active
fetus
or
mother.
Fetal
arrhythmia,
hiccups.
Connector
not
properly
connected
to
monitor.
Transducer
not
properly
placed.
Transducer
not
secured
to
patient.
Defective
transducer
cable.
No
maternal
contractions.
UA
Reference
range
exceeded.
Transducer
defective.
SECTION
7
SOLUTION
Connect
power
cord
to
receptacle.
Replace
power
cord.
Use
different
outlet.
Connect power
cord
to
monitor.
Call
Biomedical
Department
or
service
representative.
Call
Biomedical
Department
or
service
representative.
Depress
Record
pushbutton.
Install
chart
paper.
Call
Biomedical
Department
or
service
representative.
Check
that
connector
is
firmly
attached
to
monitor.
Reposition
transducer.
Apply
more
gel.
Replace
transducer.
Wait
before
moving
transducer,
FHR
will
often
return.
Use
alternate
technique.
Call
Biomedical
Department
or
service
representative.
Check
that
connector
is
firmly
attached
to
monitor.
Reposition
transducer.
Secure
transducer
to
patient.
Replace
transducer.
Remove
transducer.
Depress
UA
reference
button.
Re-apply
transducer.
Depress
UA
reference
button
again,
between
contractions.
Replace
transducer.
Call
Biomedical
Department
or
service
representative.
Troubleshooting
+
23