COSMOS ALCOHOL TESTER TYPE 2000
Software details
INSTRUCTIONS Page 08 / 2005-04-02 / TYPE 2000
The COSMOS ALCOHOL TESTER has 6 programs you can operate and adjust.
BLI Blink program 1 or 2 TON Music on or off
CAL See page 9 for Calibration LA Lamps light in 4 sectors
COU No. of tests SEN Sensitivity + seconds before display
Calling up and modifying programs
The mother board has 4 keys. + - O K and Test
You can switch from program to program using the O K, + and - keys.
The "TEST" key is only for checking measurements after calibration.
SWITCHING Key Code Program
SEQUENCE used displayed setting
Ready status - - -
1
st
setting O K END
2
nd
setting + BLI Press OK to switch to L 1 or L 2
3
rd
setting + CAL See "Calibration" on page 9
Press O K and key in, for instance, 0.58.
Then use + or – to modify the value and
press O K. The triangle will flash and the
machine is ready to be calibrated. Now
blow gas from the gas calibrator into the
machine. CAL is displayed automatically
when everything is ready.
4
th
setting + COU Press OK to call up the number of tests –
first the thousands, then the hundreds
5
th
setting + Ton Press OK for music off or on
6
th
setting + LA Press OK to check the lamps.
7
th
setting + SEN 1
st
No. = sensor sensitivity
2
nd
No. = seconds before display.
8
th
setting + END
9
th
setting O K - - - The machine is ready for use