MNL-00007 [3] Proprietary and Confidential Page 35 of 63
not be closer to the lamps than this line. Standing closer to the lamps will cause localized
streaking or erythema. Greater distances will increase the time necessary to receive a therapeutic
dose.
22.4 7 Series Lamp Removal and Replacement
Lamps may need to be replaced due to burnout or because their energy output has decreased to
the point that you may no longer be receiving an effective treatment.
DO NOT REPLACE LAMPS INDIVIDUALLY. ALL LAMPS SHOULD BE REPLACED
AT THE SAME TIME TO AVOID UNEVEN TREATMENTS OR ERYTHEMA.
PLEASE CONTACT THE DAAVLIN SERVICE DEPARTMENT PRIOR TO
REPLACING YOUR LAMPS TO OBTAIN IMPORTANT INFORMATION ON HOW
THE OUTPUT OF YOUR UNIT WILL BE AFFECTED BY THE CHANGE.
Because of differences between brands, replace lamps with the same brand as originally
installed.
1. Unplug the machine. Then, using a ¼” hex driver/socket, remove the screw from the
bottom of the grid that is in front of the lamp(s) you want to remove.
2. Lift the grid out of the holes in the bottom lamp plate, pull the bottom of the grid toward
you, then lower the grid out of the holes in the top of the unit and set it aside.
3. Grasp the lamp to be removed with both hands and press down until it clears the top
socket, then remove the lamp. Reverse the process to install lamps and grid.
Note: Please see section 10.0 Lamp Inspection on page 12, for the correct specifications of
replacement lamps and the Maintenance section for instructions on how to reset the lamp age
monitor.
23.0 M Series X and M Series T Information
23.1 Electrical Requirements
Daavlin offers the M Series Phototherapy equipment in a variety of combinations. Units contain
10 lamps and can be equipped with various combinations of UVA, Narrowband UVB, and
Broadband UVB. All domestic 110/120-volt 60 hertz models require a grounded, three prong
electrical outlet rated for at least 15 amps and comes equipped with a grounded, three prong
plug. Units designed for other voltages may be shipped with no wall plug, which would then
have to be locally obtained, and require an appropriate outlet that must be installed to meet local
standards and requirements. All power cords must be an approved hospital grade cord rated
300V, 1.5mm
2
minimum. The power cord is the main disconnect device. Each unit is equipped
with an IEC 320 style power inlet, which incorporates two, 5x20mm fuses.
Note: This unit should be on a dedicated circuit.