No Alarm Code Description Actions
Indicate the set
time is over.
- Take the follow-up actions in accordance
with the instructions from a doctor.
- Finish alarm rings after the set time.
(However, the device is not turned off
and cuffs are not deflated.)
- Press the Deflate button to deflate the
cuffs.
A certain pressure
is detected, but
pressure doesn’t
reach to the set
value or a cuff
is continuously
deflated.
- Check the connection of a cuff and a
hose.
- Check air leakage of a cuff and a hose.
- If the error persists even checking a cuff
and a hose, contact your vendor.
A cuff is not
connected to the
device.
- Check the connection of a cuff and a
hose.
- If the error persists even checking the
connection of a cuff and a hose, contact
your vendor.
Voltage error while
using the battery.
- Charge the battery by supplying AC
power.
- When the battery is used continuously
without AC power supply, it may reduce
the life span of device due to battery
discharge.
Errors during LOP
operation.
- Check the connection of LOP Sensor.
- If the error persists even checking LOP
connection, contact your vendor.
Troubleshooting
6-1. Alarm messages
* When an alarm occurs, Alarm LED turns red.
Press the Alarm button, and alarm stops. However, Low Battery Alarm keeps ringing and Time
Over Alarm is activated 30 seconds after the first alarm. Alarms occurred are displayed in the
numbers as shown in the table above on the top of LCD. When the errors occur in both 1CH
and 2CH at the same time, more urgent error is displayed.