EasyManua.ls Logo

DeFelsko PosiTector Series - Preliminary Operations; Visual Inspection; Gage Reset

DeFelsko PosiTector Series
6 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Management Procedure 2585 Rev. F
3
Table 2-2 Minimum Use Specification
Model Range Accuracy
UTG, UTG-STD, UTG-C
UTG-CA, UTG-CLF, UTG-CX
1.0 - 125 mm
(0.040 - 5.000”)
± 0.008 mm
(± 0.00025”)
UTG-M & UTG-ME
2.5 - 125 mm
(0.100 - 5.000”)
± 0.008 mm
(± 0.00025”)
UTG-P
0.200 – 12.000 mm
(0.008 – 0.472”)
± 0.003 mm
(± 0.0001”)
Table 2-3 Actual Equipment Specification
UUC Model Equipment
Generic Name
Actual Equipment Specifications Manufacturer &
Model Applicable
Range Accuracy
UTG, UTG-STD, UTG-C
UTG-CA, UTG-CLF,
UTG-CX, UTG-M &
UTG-ME
Step Blocks
2.50 - 12.50 mm
(0.100 - 0.500”)
± 0.005 mm
(± 0.00020”)
PH Tool Custom
5 Step Block
UTG-P Step Blocks
0.50 - 2.50 mm
(0.020 – 0.098”)
± 0.005 mm
(± 0.00020”)
PH Tool Custom
5 Step Block
Caution: The instructions in this Calibration Procedure relate specifically to the equipment and
conditions listed in this section. If other equipment is substituted, the information and
instructions must be interpreted accordingly.
Table 2-4 Calibration Environmental and Warm Up Requirements
Measurement Standards & Support Equipment
Environmental Requirements:
Temperature: 23 ± 5° C.
Relative Humidity: Less than 95%
Measurement Standards & Support Equipment
Warm-up and Stabilization Requirements:
None
3 Preliminary Operations
Note: Review the entire document before starting the calibration process.
3.1 Visual Inspection
3.1.1 Visually inspect the UUC for:
Probe tip wear or damage
Other damage or wear
Proper identification
3.1.2 Damage or excess wear shall be repaired prior to beginning the calibration process.
3.2 Gage Reset

Related product manuals