5
Perform a more thorough Hard Reset as follows:
Power down the Gage and wait 5 seconds.1
Simultaneously press and hold the (+) and center2
buttons until the Reset symbol appears.
This returns the Gage to a known, “out-of-the-box” condition.
It performs the same function as a menu Reset with the addition of:
Date, Time and WiFi settings are not affected by either Reset.
Bluetooth Smart = OFF
Test Time = 1 second
Sound = ON
Backlight = Normal
Flip Display = Normal
Auto Sync .net = ON
USB Drive= ON
Language = English
Battery Type = Alkaline
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following status:
NOTE:
Auto Ign. 20/90
Test Time
The PosiTector SHD has an on-screen timer to measure the
hardness of a material after a given period of time as defined by
international standards, typically 1, 3 or 15 seconds.
Use the (-) and (+) buttons to adjust the test duration. When
taking a reading, the timer starts automatically once the presser
foot is on the surface and the symbol appears on the display.
According to most international standards, readings below 20
and above 90 are not considered reliable. When enabled,
measurement results outside 20/90 will emit a low tone and will
not be included in Memory or Statistics.
If values above 90 Shore A are measured, the
PosiTector SHD-D probe should be used. If values below 20
Shore D are measured, use the PosiTector SHD-A probe.
NOTE:
(Auto Ignore)