No. Name Range Default Description/optionsNo. Name Range Default Description/options
0 = no acidic cleaning0 = no acidic cleaning
1 = one alkaline, one acidic cleaning1 = one alkaline, one acidic cleaning
2 = two alkaline, one acidic cleaning2 = two alkaline, one acidic cleaning
3 = etc.3 = etc.
Maintenance an calibration
The following points are important to sustain
correct dosing of etergents. Note that
these two tasks should be performed by an
authorised DeLaval service engineer.
• The tubes in the peristaltic pumps should
be replaced at least two times a year.
• The dosing volume (P6-P8) of the
detergent pumps should be checked
every third month. The dosing capacity
(P32-P34) may need to be adjusted.
Starting cleaning or rinsing
from the touch screen
Note: Any ongoing backflush process must
first be finished before cleaning or rinsing
can be started.
1. Ensure that the milking station is in
manual mode.
2. Remove any cow that might be in the
milking station.
After treatment window
Note: Both gates should be closed to
prevent a cow from entering before the
system cleaning has been started. Closing
and opening gates is done in the Stall
control window.
3. Press the tab After Treatment on the
touch screen to display the After treatment
window.
4. In the After treatment window, press the
Cleaning and Rinsing button. This will open
the Cleaning and rinsing window.
n Only authorised DeLaval service technicians
can repair the system. To prevent injury and
damage to the system, nothing should be
repaired by the operator.
n Do not place anything on top of the Herd
Navigator 100 device.
6 Consumables
n Consumables must be stored according to
the instructions issued. Incorrect storage or
handling may ruin them.
n Keep the cassettes in a refrigerator.
n Consumables must be handled free from dirt
during maintenance.
n Make sure the outer surfaces of the equip-
ment are clean when loading consumables.
Dirt and feed residues may attract insects
and bacteria.
7 Compliance with national regu-
lations
7.1 Compliance Statement
This equipment was tested and found to comply
with the limits for a Class B digital device, pur-
suant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection
against harmful interference in a residential
installation. This equipment generates, uses and
can radiate radio frequency energy and, if not
installed and used in accordance with the instruc-
tions, may cause harmful interference to radio
communications. However, there is no guarantee
that interference will not occur in a particular
installation.
If this equipment does cause harmful interfer-
ence to radio or television reception, which can
be determined by turning the equipment off and
on, the user is encouraged to try to correct the
interference by one or more of the following
measures:
n Reorient or relocate the receiving antenna.
n Increase the separation between the equip-
ment and receiver.