Preliminary
941322_VMS Instruction Book 2006.pdf2009-09-07
No. Name Range Default Description/optionsNo. Name Range Default Description/options
0 = no acidic cleaning0 = no acidic cleaning
1 = one alkaline, one acidic cleaning1 = one alkaline, one acidic cleaning
2 = two alkaline, one acidic cleaning2 = two alkaline, one acidic cleaning
3 = etc.3 = etc.
Maintenance an calibration
The following points are important to sustain
correct dosing of etergents. Note that
these two tasks should be performed by an
authorised DeLaval service engineer.
• The tubes in the peristaltic pumps should
be replaced at least two times a year.
• The dosing volume (P6-P8) of the
detergent pumps should be checked
every third month. The dosing capacity
(P32-P34) may need to be adjusted.
Starting cleaning or rinsing
from the touch screen
Note: Any ongoing backflush process must
first be finished before cleaning or rinsing
can be started.
1. Ensure that the milking station is in
manual mode.
2. Remove any cow that might be in the
milking station.
After treatment window
Note: Both gates should be closed to
prevent a cow from entering before the
system cleaning has been started. Closing
and opening gates is done in the Stall
control window.
3. Press the tab After Treatment on the
touch screen to display the After treatment
window.
4. In the After treatment window, press the
Cleaning and Rinsing button. This will open
the Cleaning and rinsing window.
Prohibited!
Do not use inadequate parts or
consumables.
Using products which do not meet
specified requirements, for example
spare parts or consumables, or not
appropriately trained personnel for
the DeLaval product may lead to
risks or damage. Consequently it
may also void or limit the warranty.
Caution!
Falling hazard
Ensure that the ladder is properly
secured when working at a consid-
erable height.
Warning!
Do not use spray cans containing
insecticides around the milking sta-
tion. This will damage the parts that
are made of plastic such as the tubes
for compressed-air.
5 Safe and healthy work routines
n The system requires a daily inspection to
make sure it works correctly. The inspection
is performed by the Farm Surveillance Mon-
itor (FSM) application, installed on the PC of
the farm. The customer should respond to
messages from the FSM.
n Reproduction events (calvings, insemina-
tions, pregnancy checks, dry-offs, etc.) must
be reported daily in the herd management
software.
n Use only DeLaval Herd Navigator™ 100 cas-
settes and diluent containers.
n To prevent malfunction, never load damaged
cassettes into the device.
n Always contact a local service technician if
the system stops working.
DeLaval Herd Navigator™ precision analytics 100
Safety precautions
© DeLaval 2021.
2021-05-10, Version 3 9 (44) 2800008901