OPERATING INSTRUCTIONS
39 OPERATING INSTRUCTIONS
Remarks and Appropriate
Actions
• The unit’s elapsed timer,
which advances when the
cuff is inflated, has reached
the time set-point and
caused the time alarm to
activate
• The physician should be
warned of the time up
condition. Only on the
direction of the physician,
should the time set-point be
set to a new value
• Touch the TIME indicator
and increase the time set-
point, OR
• Deflate the cuff (Refer to
Section 7 of Operation
under Section 2 Operating
Instructions)
• The battery voltage has
dropped to a level indicating
that the unit should be
plugged into AC power for
continued use, and to
charge the battery
• Plug unit in. If the unit is not
plugged in, a battery failure
condition will occur and the
unit will shut down in a ‘safe
state’ mode* opening all
valves, OR
• Press ON/STANDBY button
to shut unit down
WARNING: While running with
a Low Battery Voltage Alarm
Condition other alarm
conditions cannot be
guaranteed