6 Troubleshooting EN
11
MIDWEST
®
AUTOMATE
®
Automated Handpiece Maintenance System
6 Troubleshooting
Fault
Possible cause
Remedy
LED flashing
Spray can failure
Re-insert or replace spray can
Compressed air not connected
or not on/flowing
Turn on / connect compressed air
Door cannot be closed properly
Handpieces placed incorrectly
Handpiece chucks must be directed towards
the front or back
LED is off
No power supply
Plug in power cord
Turn unit on
Handpiece maintenance cycle
does not start
Front door not closed
Close front door
Top of oil can is more than half
filled with oil
O-Ring worn or damaged
Contact service (chapter 11)
No oil at handpiece chuck after
maintenance cycle
Clogged handpiece
Consult handpiece instructions
Oil system malfunction
Contact service (chapter 11)