3 Installation EN
7
MIDWEST
®
AUTOMATE
®
Automated Handpiece Maintenance System
3 Installation
Location
The Midwest
®
Automate
®
System may only be used in dental offices, dental hospitals and dental universities in the
ambient conditions given in the technical data chapter 10 "Specifications".
Not in intensive care units
The recommended operating temperatures are 41104°F (540°C) with humidity of 2585%
Avoid installing the unit in direct sunlight or close to a heat source (e.g. vents or radiators)
Frequency of use: up to several times a day
Length of use: from one minute to several hours a day
May be used with other units (sterilization unit, washer, etc.)
To be operated in ambient light between 100 Lux and 1,500 Lux
Ambient temperature in immediate proximity of the unit during operation: max. 35°C (95°F)
Unit should be placed on the soiled side of the sterilization area.
The unit should be operated in a clean area
The unit must be operated in well ventilated areas to avoid possible explosive conditions
The unit should be placed on a flat, level, water-resistant surface
Never install and operate the unit on a sloped surface. Surface must be adequate to accommodate the front door when
it is fully open.
Electromagnetic environment
The unit has been tested and meets applicable standards for electromagnetic emissions. While the unit does not emit
any radiation, it may itself be affected by other equipment which does. We recommend that the unit be kept away from
potential sources of interference.
Electrical requirements
Use properly grounded and fused power sources with the same voltage rating as indicated on the label at the
back of the unit. Avoid multiple outlet receptacles. If using a surge suppressor power bar, plug in one unit only.
The power input plug on the back of the unit is to be used as a separating device to the power supply network.
Power and air connections
Insert power plug into socket.
Fasten and secure compressed air hose onto coupling.
Ensure compressed air regulator indicates 4 bar (58 psi).
To adjust inlet air pressure, pull out handle on the compressed air regulator and turn clockwise (increase) or counter-
clockwise (decrease) until 4 bar (58 psi) is displayed.
Push handle on the compressed air regulator back in and lock it.
Use compressed air according to the information contained in the chapter 10 "Specifications", to ensure proper function.
Installing spray can
(see chapter 5 "Maintenance")
Attaching maintenance adaptors to handpiece cleaning ports
(see chapter 4 "Operation")

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dentsply Sirona MIDWEST AUTOMATE and is the answer not in the manual?

Dentsply Sirona MIDWEST AUTOMATE Specifications

General IconGeneral
BrandDentsply Sirona
ModelMIDWEST AUTOMATE
CategoryDental equipment
LanguageEnglish

Related product manuals