EasyManua.ls Logo

DENTSPLY Cavitron Steri-Mate - User Manual

DENTSPLY Cavitron Steri-Mate
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Directions for Use
Mode D’Emploi
Instrucciones de Uso
Gebrauchsanleitung
Istruzioni per L’Uso
Указания по применению
Please read carefully and completely before use.
Veuillez lire attentivement le mode d’emploi suivant avant toute utilisation.
Lea atentamente toda la información antes de usar el producto
Bitte vor Inbetriebnahme der Einheit sorgfältig und vollständig durchlesen.
Leggere attentamente le istruzioni prima dell’uso.
Прочтите внимательно все указания перед началом работы.
Cavitron
®
Handpiece
Sterile Lavage Kit
Pièce à main Cavitron
®
Steri-Mate
®
Kit de lavage stérile
Kit de lavado estéril para Pieza de
mano Cavitron
®
Steri-Mate
®
Steriles Spülkit für
Cavitron
®
Steri-Mate
®
Handstück
Manipolo Cavitron
®
Steri-Mate
®
Kit di lavaggio sterile
Наконечник Cavitron
®
Steri-Mate
®
Комплект для стерильного орошения

Table of Contents

Questions and Answers

  • R
    Rebecca LeeAug 13, 2025
    What to do if DENTSPLY Cavitron Steri-Mate has lavage leakage?
    • G
      Gregory PowellAug 13, 2025
      If you notice lavage leakage from your DENTSPLY Dental equipment, several factors could be the cause. Start by checking the lavage solution bag for punctures and replace it if damaged. Inspect the lavage tubing assembly for porosity and replace if necessary. Examine the handpiece adapter and Steri-Mate® Handpiece for cracks, replacing them if found. Ensure the tubing size is correct and fits snugly on the adapter barb; replace with .110 inch / 2.794 mm I.D. tubing if it's loose. Inspect the O-rings on the Cavitron® Insert, handpiece cable, and handpiece adapter for wear or if they're missing, and replace them as needed using the appropriate item numbers (Green Item #62605, Black Item #62351, and Item #79357).
  • S
    Stephanie SchmidtAug 25, 2025
    Why DENTSPLY Cavitron® Insert does not vibrate?
    • R
      rivaskristiAug 25, 2025
      If the Cavitron® Insert on your DENTSPLY Dental equipment isn't vibrating, the issue could stem from a few components. Try replacing the insert with a new Cavitron® Insert to rule out a defective insert. If that doesn't solve the problem, the Steri-Mate Handpiece might be defective; replace it with a known working handpiece. Alternatively, the Steri-Mate Handpiece Adapter could be the cause. To check, remove the Adapter and plug the Handpiece directly into the handpiece cable; if the Cavitron® Insert then scales properly, replace the Adapter.
  • S
    Shannon WilliamsNov 3, 2025
    How to fix no lavage flow from DENTSPLY Cavitron® Scaler?
    • D
      Debra JohnsonNov 3, 2025
      If you're experiencing no lavage flow from your DENTSPLY Cavitron® Scaler, it could be due to several reasons. First, ensure the sterile lavage solution bag isn't empty and replace it if needed, remembering it's for single-patient use only. Check that the bag is correctly installed, centered in the pressure infuser sleeve. Verify the pressure infuser is working by ensuring the release valve is closed and pumping the infuser bag to maintain a pressure between 50 to 150 mm Hg. Also, adjust the Flow Roller Clamp on the tubing for proper flow. Make sure the tubing Flow Roller Clamp is open to allow fluid flow. Inspect for air leaks in the gauge, inflation bulb, air bladder, or T-fitting, correcting any tubing connection leaks, and replace the Pressure Infuser assembly if necessary. Check the h...

Summary

Introduction to Cavitron Steri-Mate Handpiece

Indications for Use

Details the intended applications of the sterile lavage kit in dental treatments.

Contraindications

Specifies conditions or procedures where the kit should not be used.

Warnings and Precautions for Use

System Precautions

Outlines precautions for the proper setup and environment of the sterile lavage system.

Infection Control Procedures

General Information on Infection Control

General guidelines for infection control procedures with the kit.

Accessories and Replaceable Parts

Available Accessories

Lists available accessories for the sterile lavage kit.

Replacement Parts

Lists components available for replacement in the kit.

Procedure for Using the Lavage Kit

Troubleshooting and Support

Troubleshooting Guide

Provides solutions for common issues encountered with the lavage system.

Technical Support and Repairs

Contact information for technical assistance and repair services.

Product Warranty Information

Worldwide Service Centers

DENTSPLY Cavitron Steri-Mate Specifications

General IconGeneral
BrandDENTSPLY
ModelCavitron Steri-Mate
CategoryDental equipment
LanguageEnglish

Related product manuals