Initial out-of-box startup: Select default 8:1 handpiece; TUL-8M (Tulsa Dental Specialties) or
AHP-88MN/MNP (Anthogyr). It is important to select correct handpiece for the system to control
speed and torque accurately. An 8:1 handpiece is required for reciprocation.
Calibration Procedure: Connect motor and handpiece to system. Secure a file into the hand-
piece. Press & release CAL button. To ensure torque accuracy, calibrate system daily.
WARNING: During the preprogrammed calibration operation, the file will rotate.
Basic Operation: Select the Preset Button corresponding to the endodontic rotary file series
you are using. By default, the PROMARK
®
Preset Buttons are set for the file series shown
below. Pressing the Preset Button repeatedly will cycle through the individual files listed below
each series name. The Up/Down Arrow Buttons can also be used to cycle through the files,
and the Right/Left Arrow Buttons can be used to navigate from Preset to Preset.
System Setup: Enter the System Setup menu by holding down the CAL button until a double
beep is heard. (See Directions For Use for additional setup instructions.)
* Refer to Directions For Use for advanced features
and programming.
Distributed By
DENTSPLY Tulsa Dental Specialties
DENTSPLY International, Inc. 608 Rolling Hills Drive • Johnson City, TN 37604
1-800-662-1202 • Fax:1-800-597-2779 • www.TulsaDentalSpecialties.com
Preset Button 1 Default File Series:
DENTSPLY Tulsa Dental Specialties PathFile
®
All Files
File #1: PathFile
®
(All Files)
Preset Button 2 Default File Series:
DENTSPLY Tulsa Dental Specialties WaveOne
®
All Files
File #1: WaveOne
®
(All Files)
An 8:1 handpiece is required for reciprocation.
Preset Button 3 Default File Series:
DENTSPLY Tulsa Dental Specialties ProTaper Next™
File #1: PTN High Torque (All)
File #2: PTN Low Torque (All)
Preset Button 4 Default File Series:
DENTSPLY Tulsa Dental Specialties ProTaper
®
Universal
File #1: ProTaper S1 & SX
File #2: ProTaper S2 & F1
File #3: ProTaper F2, F3, F4, F5
File #4: ProTaper D1 & D2
File #5: ProTaper D3
Preset Button 5 Default File Series:
DENTSPLY Tulsa Dental Specialties VORTEX
®
04, 06 Taper
File #1: VORTEX Orifice (All)
File #1: VORTEX 04, .35 through .50 taper
File #2: VORTEX 04, .25 and .30 taper
File #3: VORTEX 04, .15 and .20 taper
File #4: VORTEX 06, .35 through .50 taper
File #5: VORTEX 06, .25 and .30 taper
File #6: VORTEX 06, .15 and .20 taper
PROMARK
®
Default Preset Chart
Quick Reference Guide
*
Speed
Increase/Decrease
Torque
Increase/Decrease
Motor Direction
Forward/Reverse
Auto-Stop-Reverse/
MAX Torque
Reciprocation
Out Mode
Standby Key
Preset Buttons (5)
Calibration (single press)
System Setup (press & hold)
• SETUP OPTIONS:
- Preset Memory Lock
- Recall Factory Setup
• GLOBAL OPTIONS:
- Auto Stop Reverse
- Reverse Beeper
- Torque Warning
- Accessory Warn
- Sleep Mode
- Keypress Beep
- System Volume
- Save Settings
- Cancel
Menu Navigation & Select
Handpiece Ratio Selection
Question and Answer IconNeed help?

Do you have a question about the DENTSPLY PROMARK and is the answer not in the manual?

DENTSPLY PROMARK Specifications

General IconGeneral
BrandDENTSPLY
ModelPROMARK
CategoryDental equipment
LanguageEnglish

Summary

PROMARK Device Overview and Operation

PROMARK Controls and Navigation

Details the physical controls, menu navigation, and key buttons for operating the PROMARK system.

Initial System Setup

Guides users through the initial out-of-box startup process, including handpiece selection.

Calibration Procedure

Explains the steps for calibrating the system to ensure accurate torque, recommending daily calibration.

Basic Operation Guide

Instructions on how to select endodontic rotary file series using preset buttons and navigation keys.

PROMARK Default Preset Chart

Preset Button 1: PathFile®

Default file series settings for Preset Button 1, including PathFile® all files.

Preset Button 2: WaveOne®

Default file series settings for Preset Button 2, including WaveOne® all files.

Preset Button 3: ProTaper Next™

Default file series settings for Preset Button 3, including ProTaper Next™ High and Low Torque.

Preset Button 4: ProTaper® Universal

Default file series settings for Preset Button 4, covering ProTaper® Universal files S1-SX to D3.

Preset Button 5: VORTEX® 04, 06 Taper

Default file series settings for Preset Button 5, including VORTEX® Orifice and 04/06 Taper files.