26. Technical Specification
Blue (415nm)
- Maximum optical power is 10 j/cm2
- Maximum operating time per treatment is 30 minutes
Red (633nm)
- Maximum optical power is 20 j/cm2
- Maximum operating time per treatment is 30 minutes
NIR (830nm)
- Maximum optical power is 10 j/cm2
- Maximum operating time per treatment is 30 minutes
Maximum variation of 10% optically
Ocular hazard distance (EN 60601-2-57): 0.025m when in use with the Base unit,
0.0m when directly placed on the body
EN 62471 Risk Group 2
Mains supply: 100 -240 Vac, 50 / 60 Hz, 1A.
PSU output, 30Vdc, 1.2A 36W
Tailor to clients with 7 treatment options.
Weight: 2.5kg
Footprint (in Base Unit) W520 x L335 x H300 mm
Footprint (Array in flat position): W280 x L790 x H30 mm
Duty Cycle: Continuous
IP Rating : IP20
26.1 Electromagnetic Guidance
Emissions Test Compliance
RF conducted and radiated emissions CISPR 11 Group 1, Class B
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3 Complies
Immunity tests Immunity Test Level
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Proximity fields from RF
wireless communications equipment
IEC 61000-4-3
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 27 V/m
Electrical fast transients/burst
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Surges
IEC 61000-4-5
Line to line: ± 0,5 kV, ± 1 kV
Conducted disturbances induced by RF fields
IEC 61000-4-6
3 V/m
0,15 MHz – 80 MHz
6 V/m in the ISM and Amateur Radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Voltage dips and interruptions
IEC 61000-4-11
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle at 0°
70 % UT; 30 cycles at 0°
0 % UT; 300 cycles
UT is the AC mains voltage prior to application of
the test level.
Table 11: Electromagnetic Technical Information.