3.2 Operation of the PRO-II
®
and PRO-III
®
Devices
3.2.1 Operation at initial startup
1. Press the button for 1-2 seconds to switch on the PRO-II
®
or PRO-III
®
device. The
following start screen is displayed for 5 seconds:
2. At initial startup the unit will open with Continuous Mode set to -80mmHg:
3. Use the arrow buttons to select the Language menu.
4. Use the button to confirm your choice.
5. Use the arrow buttons to select the desired language:
6. Use the button to confirm your choice.
7. Select the prescribed operating mode in the Setup menu.
8. Use the button to confirm your choice.
9. Perform the desired treatment (please refer to chapter 3.2.2 or 3.2.3).
10. Switch off the PRO-II
®
or PRO-III
®
device by pressing the button for 3 seconds.
DeRoyal
23
3.2.2 Continuous operation
In the continuous operating mode, the PRO-II
®
and PRO-III
®
devices
maintain a continuous
vacuum in the wound dressing.
The vacuum pressure value can be set from -20 mmHg to -200 mmHg in increments of
5 mmHg. The default value is -80 mmHg.
1. Press the button for 1-2 seconds to switch on the PRO-II
®
or PRO-III
®
. The
following start screen is displayed for 5 seconds:
2. While the start screen is displayed, simultaneously press the arrow buttons. The
menu Setup is displayed.
3. Use the arrow buttons to select the Continuous menu.
4. Use the button to confirm your choice.
The following screen is displayed:
5. Use the arrow buttons to set the prescribed vacuum value.
6. Confirm the setting by pressing the button.
The following overview screen is displayed:
7. Press the button to start the therapy.
(Default target value: -80 mmHg)
8. The following screen is displayed for one minute after the therapy has been started:
The bar symbolizes the tightness of the wound dressing. If the bar is in the too tight”
or not tight” range, the wound dressing must be adjusted or may need to be
reapplied. The wound dressing is properly applied if the bar is in the middle range.