36
7 Technical Data
7.1 PRO-II
®
Device
Flow rate
Max. 8 l/min
Pressure
Max. -200 mmHg;
Conversion factor: 1 kPa ~ 7.5 mmHg
Canister
Disposable exudate canister (250 ml or 450 ml)
Suction tube
Suction tube with silicone tubing connector ~ length
150 cm
Power supply unit
FRIWO FW 7555M/12 or DeRoyal supplied power
adapter
Nominal voltage of the power
supply unit
In: AC 100 240 V~ / 50-60 Hz / 350 150 mA
In (UL only): 120 Vac / max. 350 mA / 50-60 Hz
Out: DC 12 V / 1,25 A
Maximum load current at 12 V
1.25 A
Nominal voltage of the circuit
board at 12 V
12 V
Power consumption
15 W
Permissible input voltage
1.25 A
Protection class as per
IEC 60601-1
Type BF
Protection class as per
IEC 60601-1
II
Degree of protection (IP code) as
per IEC 60529
IP22
UL marking
Sound emission
Operation:
35 dB (A)
High priority alarm:
52 dB (A)
Low priority alarm:
49 dB (A)
Ambient conditions
Transport/Storage:
-25°C to +60°C
humidity of max. 93%
non-condensing
Operation:
+5°C to +40°C
humidity 15% to 93%
non-condensing
Air Pressure:
700 hPa to 1060 hPa
Battery, rechargeable
7.4 V, 4.4 Ah Lithium ions
Charging time if battery is empty
Charging time if battery is
approx. 50% full
6 - 7 hours
3 - 3.5 hours
Dimensions (H x W x D) in mm
165 x 220 x 90
Weight (base unit)
1.2 kg
Pressure measurement accuracy
Target pressure > -80 mmHg
max. 5 %
Target pressure < -80 mmHg
max. 10 %
Operating time
Continuous operation
Battery-powered runtime
approx. 24 48 hours, depending on the run time of
the motor
Item number
NP-2000; NP-2000GPS
AS TO ELECTRICAL
SHOCK, FIRE AND
MECHANICAL HAZARDS
ONLY IN ACCORDANCE
WITH ANSI/AAMI
ES60601-1 (2005),
CAN/CSA-C22.2 No.
60601-1 (2008)
37
7.2 PRO-III
®
Device
Flow rate
Max. 8 l/min
Pressure
Max. -200 mmHg;
Conversion factor: 1 kPa ~ 7.5 mmHg
Canister
Disposable exudate canister system (800 cc)
Power supply unit
FRIWO FW 7555M/12 or DeRoyal supplied power
adapter
Nominal voltage of the power
supply unit
In: AC 100 240 V~ / 50-60 Hz / 350 150 mA
In (UL only): 120 Vac / max. 350 mA / 50-60 Hz
Out: DC 12 V / 1,25 A
Maximum load current at 12 V
1.25 A
Nominal voltage of the circuit
board at 12 V
12 V
Power consumption
15 W
Permissible input voltage
1.25 A
Protection class as per
IEC 60601-1
Type BF
Protection class as per
IEC 60601-1
II
Degree of protection (IP code) as
per IEC 60529
IP22
UL marking
Sound emission
Operation:
35 dB (A)
High priority alarm:
53 dB (A)
Low priority alarm:
51 dB (A)
Ambient conditions
Transport/Storage:
-2C to +60°C
humidity of max. 93%
non-condensing
Operation:
+5°C to +40°C
humidity 15% to 93%
non-condensing
Air Pressure:
700 hPa to 1060 hPa
Battery, rechargeable
7.4 V, 4.4 Ah Lithium ions
Charging time if battery is empty
Charging time if battery is
approx. 50% full
6 - 7 hours
3 - 3.5 hours
Dimensions (H x W x D) in mm
290 x 259 + 100 (canister) x 130
Weight (base unit)
2.2 kg
Pressure measurement accuracy
Target pressure > -80 mmHg
max. 5%
Target pressure < -80 mmHg
max. 10%
Operating time
Continuous operation
Battery-powered runtime
approx. 24 48 hours, depending on the run time
of the motor
Item number
NP-3000
AS TO ELECTRICAL
SHOCK, FIRE AND
MECHANICAL HAZARDS
ONLY IN ACCORDANCE
WITH ANSI/AAMI
ES60601-1 (2005),

Table of Contents