EasyManua.ls Logo

Det-Tronics 505 - Page 9

Det-Tronics 505
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Use formula: C x K = S
C = Concentration of calibration gas in % LFL = 50
K = K-factor from Table 1 in GTN01 = 1.44
50 x 1.44 = 72
S = 72
2. Use formula: (S x 0.0067) + 0.17 = S1
S = span output level (determined in step 1) = 72
(72 x 0.0067) + 0.17 = S1
(72 x 0.0067) = 0.48
0.48 + 0.17 = 0.65
S1 = 0.65
3. Calibrate the Model 505 for a reading of 0.65 vdc
on the voltmeter with a calibration mixture of 50%
LFL methane applied to the sensor.
NOTE
This procedure applies only to the Model 505
Transmitter. Other Det-Tronics transmitters use
the standard K-factor formula as described in
GTN01.
7 95-8472
WARNING
Before removing the junction box cover, verify that no dangerous levels of gas are present.
Step Switch Position Operator Action
1 CAL/NORM switch in the CAL position. 1. LED turns on.
2. Connect a digital voltmeter to the transmitter test jacks.
3. Set the meter range to 2 vdc.
2 ZERO/SPAN switch in the ZERO position. 1. Adjust the ZERO potentiometer to read 0.000 vdc on the voltmeter.
See Note 3 below.
3 ZERO/SPAN switch in the SPAN position. 1. Adjust the 4 ma potentiometer to read 0.167 vdc on the voltmeter.
2. Apply the 50% LFL calibration gas to the sensor. When the output has
stabilized, adjust the SPAN potentiometer for a reading of 0.500 on the
voltmeter.
4 ZERO/SPAN switch in the ZERO position. 1. Sensitivity test. The meter must read greater than 0.015 vdc. See Note
4 below.
2. Remove the calibration gas.
3. When the meter reads 0.002 vdc or less, remove the test probes.
5 CAL/NORM switch in NORM position. 1. The LED turns off.
2. The calibration is complete.
3. Replace the junction box cover.
NOTES:
1. When the CAL/NORM switch is in the CAL position, the yellow LED turns on and the 4 to 20 ma output signal goes to 3.4 ma.
2. The voltmeter must be suitable for use in a hazardous location.
3. If the possibility of background gases exists, purge the sensor with clean air prior to the zero adjustment to assure accurate
calibration.
4. A typical sensitivity reading with 50% LFL gas applied to the sensor is 35 to 50 millivolts for a new sensor. Sensor
replacement is recommended when the sensitivity reading is less than 15 millivolts.
5. If a dust cover or splash shield is used, inspect it to be sure that it is not dirty or plugged. A plugged dust cover can restrict the
flow of gas to the sensing element, seriously reducing its effectiveness. For optimum performance, sensor covers/filters
should be replaced frequently to ensure that they are not degraded or plugged.
Table 2—
Calibration Procedure