8529-0637-2M Rev C DL Series Owner’s Manual
7
PRINTER, CONT.
Step 6 Perform Gap Calibration
1. Turn the scale OFF.
2. Press and hold the Feed key, then turn the scale ON.
3. Continue to hold the Feed key until the self-checking is finished, and the scale goes into
normal operation mode.
3a. The printer will roll out the paper at a length of approximately 50mm.
3b. After the paper is rolled out, the problem should have been eliminated. If there is an
error code (E7.10), it means there is still a gap calibration issue. Please try the
procedure again after confirming the correct paper has been used.
4. Press the Feed key to print the test page.
5. Turn the scale OFF.
Step 7Replace Printer Cover Door
1. Replace the printer cover door and re-install the weighing platform.
2. The scale is ready for normal operation mode.
Replace Printer Cover Door