Page | 16
Editing Logger/Repeater Settings
Once the system has been setup existing Loggers & Repeaters can be modified. All fields can be modified
except for Unit ID. There are two options for changing Logger/Repeater setup:
Making changes wirelessly:
o Logger must be actively communicating – no lost communication errors and time of last
transmission should have been updated within the past 7 minutes. If the logger is not
communicating change via USB connection.
o How to Edit:
From the Logger tab (Loggers only), Structure tab or Devices tab right click on the
Logger/Repeater and select edit to open Edit Logger window.
Modify field setting. More than one field can be changed at a time.
Click on SAVE CHANGES AND START. A window will pop up indicating that the
change has been sent via Wireless mode. Leave the Logger in its current location until
the wireless change is complete.
The unit can not be edited and the Edit option will remain grayed out until the changes
have been completed. A “% Change Complete” field will pop up to the left of the logger
name field to indicate wireless change status.
The receiver will try 3x to send changes. If the changes are not sent successfully, an
error message will appear indicating that the wireless change has failed. Bring the unit
to the PC and edit the unit via USB.
o The following changes can NOT be made wirelessly (logger must be connected to the PC to
make changes:
Connect Via
Unit ID
Push to Start
Making a logger active or inactive
Status
USB Change: Use this option in the event that the Logger has lost communications or you are
making a change that cannot be made wirelessly
Connect the Logger to the PC via the USB cable.
The Edit Logger window will open.
Modify field(s). More than one field can be changed at a time.
Click on SAVE CHANGES AND START. The Updating Logger window will open.
Once the update is complete, unplug the USB and place the Logger/Repeater in the desired
logging location.