EasyManuals Logo

Dini Argeo HLD3000 User Manual

Dini Argeo HLD3000
13 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
12
CALIBRATION WEIGHTS
MODEL
HLD150
HLD300
HLD600
HLD1500
HLD3000
HLD6000
1° SAMPLE WEIGHT
50g
100g
200g
500g
1000g
2000g
2° SAMPLE WEIGHT
100g
200g
400g
1000g
2000g
4000g
3° SAMPLE WEIGHT
150g
300g
600g
1500g
3000g
6000g
P2 CAL CALIBRATION WITH FREE SAMPLE WEIGHT
By pressing ZERO it’s possible to access the calibration with free sample weight.
- The display shows “UNLOAD”;
- Empty the platter, after Stable indicator on the weight will be acquired automatically;
- Set the value of the mass weight by using the MODE key to clear the value, the TARE key to increase the value
of the digit, the PCS key to go on the previous digit and the % key to go on the following digit.
- Confirm the entered value with ZERO; the display shows “LOAD”;
- Put the mass on platter, after stable indicator on the weight will be acquired automatically.
- The indicator returns to weighing.
P3 Cnt A/D CONVERTER POINTS
By pressing ZERO the instrument shows the A/D converter points relative to the weight on the scale.
Press PRINT/M+ to exit the step.
P4 AZ ZERO TRACKING
By pressing ZERO one sets the divisions' number for the zero tracking, in other words, scale’s thermal drift
compensation parameter; the set value corresponds to the number of divisions which is cleared.
Press MODE to select one of the suggested divisions: oFF, 0.5d, 1d, 2d, 4d.
P5 GrA GRAVITY COEFFICIENT
Trough this step one access the acceleration value setting of instrument calibration and use zone.
Manual entry of the g value: Set the value by using the MODE key to clear the value, the TARE key to increase the
value of the digit, the PCS key to go on the previous digit and the % key to go on the following digit; one must modify
all the 6 digits of the gravitational acceleration. Confirm the entered value with ZERO or press PRINT/M+ to exit
the step.
P6 CAP SCALE CAPACITY
By pressing ZERO one selects the total capacity of the scale.
Press MODE to select one of the suggested divisions: 150, 300, 600, 1500, 3000, 6000 g.