EasyManua.ls Logo

Dräger Alcotest 6000

Dräger Alcotest 6000
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Disposal | en
Dräger Alcotest 6000 21
Respiratory volume >1.2 L (default)
Breath flow >10 L/min (default)
Duration of breath >1.6 s (default)
Measurement accuracy
Standard deviation of the repeatability with ethyl alcohol standard
<0.50 mg/L
0.50 mg/L
<0.100 % BAC
0.100 % BAC
0.008 mg/L or
1.7 % of the measured value
0.0017 % BAC or
1.7 % of the measured value.
Whichever value is higher
Wait time for measurement readiness following a previous mea-
surement with:
0.00 mg/L: 2 s
0.25 mg/L: 20 s
0.50 mg/L: 40 s
1.00 mg/L: 80 s
>1.50 mg/L: 120 s
0.00 %: 2 s
0.050 %: 20 s
0.100 %: 40 s
0.200 %: 80 s
>0.300 %: 120 s
Test for accuracy
Recommendation: every 12 months (local deviations are possible)
Dimensions
50/60 mm x 141 mm x 31 mm (W x H x D)
1.92/2.36 in x 5.55 in x 1.22 in (W x H x D)
Weight
approx. 5.3 oz (150 g)
Power supply
1 x 3 V CR123A
The battery is designed for up to 1500 measurements (depending on
use and the ambient conditions)
Interface
Micro-USB, 2.0 fullspeed, not for charging the device, for service only
Protection class
IP 54
Interface

Related product manuals