Important Safety Information
10
Operating Instructions Caleo, Software 2.n
Precautions During Preparation
WARNING !
Always observe all precautions against fire hazards from
oxygen (see page 61).
WARNING !
Never leave infant unattended when access doors or
access ports are open to avoid any risk of an infant
falling out of an incubator.
WARNING !
Always ensure that access port catches are securely
engaged in order to avoid any risk of an infant falling out
of an incubator.
WARNING !
When opening and closing front door, avoid pinching or
jamming hoses and cables in the attached double wall.
WARNING !
Always ensure that both knobs of the large access doors
are engaged in position in order to avoid any risk of an
infant falling out of an incubator.
The red catch behind each knob must no longer be
visible!
WARNING !
The attached double wall must be positioned parallel to
the front door otherwise the ducted flow of warm air will
be interrupted!
WARNING !
When opening and closing side doors, avoid pinching or
jamming hoses and cables in the attached double wall.
WARNING !
Always ensure that both knobs of the side doors are
engaged in position in order to avoid any risk of an infant
falling out of an incubator.
The red catch behind each knob must no longer be
visible!
WARNING !
Always observe the maximum load of the infant bed
(5 kg, 11 lbs).
WARNING !
Never leave infant unattended when the bed has been
pulled out to avoid any risk of an infant falling out of an
incubator.
WARNING !
Do not lean on the bed when it is pulled out. Equipment
damage with risk of patient injury may result.
WARNING !
Always ensure that the bed is pushed all the way in!
Otherwise the ducted flow of warm air will be interrupted,
and the infant may be warmed or cooled excessively.
WARNING !
Do not use the X-ray drawer as a writing support or as a
bed for the infant. Risk of equipment damage or patient
injury.
WARNING !
Always ensure that the x-ray drawer is pushed all the
way in! Otherwise the ducted flow of warm air will be
interrupted, and the infant may be warmed or cooled
excessively.
WARNING !
Make sure that the installed hoses are routed freely and
safely.
WARNING !
Do not reach between incubator housing and housing
support while tilting the bassinet. Risk of injury!