Important Safety Information
11
Operating Instructions Caleo, Software 2.n
WARNING !
Use only authentic, originally sealed OEM bags
containing sterilized distilled or demineralized water.
Do not use any additives for water intended for
humidifying the incubator.
Do not use infusion solution bag for the humidification
system.
WARNING !
Output of the integrated power strip is not monitored!
Do not connect life support devices which do not have
their own power failure alarm.
Do not exceed the maximum permissible power input
for connected accessories
(all 4 sockets together: max. 2 A).
Do not exceed the maximum permissible total leakage
current. For the leakage current of Caleo without socket
strip see "Technical Data", page 166.
WARNING!
Installation of the nurse call kit should only be performed
by DrägerService or factory trained and authorized
service personnel.
WARNING!
Connect nurse call to a central hospital alarm system
only while Caleo is properly grounded via its power cable
and a grounded wall outlet or via the grounding pin on
the rear panel of the control unit.
Otherwise, the risk of electric shock cannot be safely
excluded.
WARNING!
Only alarm messages with a high risk potential or
serious device faults are transmitted via nurse call.
WARNING!
Connection of a nurse call does not relieve staff of their
duty to check patient data at regular intervals.
z Screen displays must be checked regularly.
WARNING!
A fault in any of the components in the link between
nurse call and central hospital alarm system (e.g. in the
electronics for nurse call in Caleo, in the Caleo power
supply,
or in the enunciator of the central hospital alarm
system, etc.) may result in failure of the nurse call.
WARNING!
The system must meet the requirements about medical
electrical equipment in accordance to IEC/EN 60601-1-1
and IEC/EN 60601-1-2.
WARNING!
Connect external devices to the interfaces only while
Caleo is properly grounded via its power cable and a
grounded wall outlet or via the grounding pin on the rear
panel of the control unit.
Otherwise, the risk of electric shock cannot be safely
excluded.
WARNING!
All data that are transmitted via medical devices are for
information only and should not be used as the sole
basis for clinical decisions.
WARNING !
Do not use a power outlet strip for supplying power to
the Caleo incubator!
Connecting the incubator via a power outlet strip may, in
case of failure of the protective earth conductor, cause
patient leakage currents to rise above permitted limits
with a risk of electric shock to the patient.
WARNING!
The Caleo incubator is ready for operation only when all
checks have been performed successfully.