Operation
106
Operating Instructions Caleo, Software 2.n
Configuration
Configuring system settings
From the configuration mode selection menu:
1 Turn and press rotary knob to configure unit of temperature.
– The temperature unit is highlighted by a light background.
1 Turn rotary knob to select unit.
1 Press rotary knob to confirm unit.
– The selected unit is displayed on screen.
Unit of weight, display of dT/T
2, central alarm indicator, eleva-
tion above sea level, and contrast are configured in the same
fashion.
– The unit of weight can only be configured if Caleo is
equipped with the integrated infant scale*.
– When using two skin temperature sensors, T
2 or the
difference between T
1 and T2 (dT) can be displayed in
addition to the value of T
1.
– If the central alarm LED is disabled, alarm situations are
only indicated by the blinking display for the measured
value, blinking bar LED on the control unit, and the audible
alarm.
– Elevation above sea level can only be set if Caleo is
equipped with integrated O
2-control. An incorrectly set
elevation will reduce the measuring accuracy of the O
2
sensors (approximately 1.5 % additional error at 3280 ft /
1000 m).
– If the screen becomes difficult to read, the contrast setting
can be adjusted.
z To cancel a selection, press » « soft key.
– The display returns to the configuration mode selection
menu (see page 104).
* available option