Operation
106
Operating Instructions Caleo, Software 2.n
Configuration
Configuring system settings
From the configuration mode selection menu:
1 Turn and press rotary knob to configure unit of temperature.
The temperature unit is highlighted by a light background.
1 Turn rotary knob to select unit.
1 Press rotary knob to confirm unit.
The selected unit is displayed on screen.
Unit of weight, display of dT/T
2, central alarm indicator, eleva-
tion above sea level, and contrast are configured in the same
fashion.
The unit of weight can only be configured if Caleo is
equipped with the integrated infant scale*.
When using two skin temperature sensors, T
2 or the
difference between T
1 and T2 (dT) can be displayed in
addition to the value of T
1.
If the central alarm LED is disabled, alarm situations are
only indicated by the blinking display for the measured
value, blinking bar LED on the control unit, and the audible
alarm.
Elevation above sea level can only be set if Caleo is
equipped with integrated O
2-control. An incorrectly set
elevation will reduce the measuring accuracy of the O
2
sensors (approximately 1.5 % additional error at 3280 ft /
1000 m).
If the screen becomes difficult to read, the contrast setting
can be adjusted.
z To cancel a selection, press » « soft key.
The display returns to the configuration mode selection
menu (see page 104).
* available option