Operation
107
Operating Instructions Caleo, Software 2.n
Configuration
Configuring alarm settings
From the configuration mode selection menu:
1 Turn and press rotary knob to configure alarms.
The menu that is then displayed allows the user to select
Kangaroo Mode alarm settings, the initial volume of audible
alarms and the alarm thresholds for temperature control and
O
2 control.
1 Turn rotary knob to select menu item.
1 Press rotary knob to confirm menu item.
z To cancel a selection, press » « soft key.
The display returns to the configuration mode selection
menu (see page 104).
Kangaroo Mode:
The text items in the description that follows are examples:
1 Select skin alarm T1 min (lower alarm limit for skin
temperature) = turn and press the rotary knob.
When selected, the setting for "alarm skin T1 min" is
highlighted by a bold frame.
1 Turn rotary knob to select alarm setting.
1 Press rotary knob to confirm alarm setting.
The display returns to the alarm settings menu (see above).
The other alarms are set using the same method.
z To cancel a selection, press » « soft key.
The screen returns to the configuration mode selection
menu (see page 104).
For a description of Kangaroo Mode alarms, please refer to
page 157.
Setpoint ranges:
Skin alarm T1 min 33
o
C to 37
o
C and off
Skin alarm T2 min 33
o
C to 37
o
C and off
dT alarm min –2
o
C to 2
o
C and off
dT alarm max 2
o
C to 5
o
C and off
Default values: last settings