Operating Instructions Caleo, Software 2.n
Disassembly
Disassembly
Checking the air intake filter
z Tilt the unit (see page 48) for easy removal of the air intake
filter.
1 Check expiration date: see label on the unit.
2 Press down the knurled part of the clip and open the filter
cover.
If a filter is already in place:
z Check the condition of the filter.
Replace filter when
– it is dirty, damaged, or moist.
– the expiration date has been reached.
– the filter has been used with an infectious patient.
z Discard old filter.
z Install a new filter.
Inserting the filter
z Insert the filter in the open filter cover.
2 Close the filter cover.
Check that the filter cover is securely closed.
z Write down the expiration date of the new filter (2 months
from date of installation) on the label.
1 Stick label to the device.