Care
122
Operating Instructions Caleo, Software 2.n
Disassembly
Removing the water supply
z Close clamp on the water supply connection kit.
z Remove and dispose of water bag and water supply
connection kit and dispose of properly.
Or:
1 Remove water container from its holder.
2 Detach water supply connection kit from the water
container.
z Dispose of the water supply connection kit.
z Either clean the water container in a parts washer at 93
o
C
(200
o
F)*,
or
z sterilize at 134
o
C (273
o
F)*.
z Tilt Caleo bassinet back to the horizontal position (see
page 48).
z Switch on cleaning mode** (see page 101).
* Observe national and international standards regarding procedures
for cleaning, disinfection, and sterilization.
** Available option