Care
128
Operating Instructions Caleo, Software 2.n
Disassembly
Control unit
z Pull rotary knob from the control unit.
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect the surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
CAUTION !
Do not allow any moisture to enter the control unit.
Do not disinfect control unit by immersion or spraying.
Equipment damage may result.
CAUTION !
Ensure that only recommended cleaning agents and
disinfectants are used!
The acrylic and Makrolon material may develop stress cracks
if other agents, such as alcohol, are used.
Do not use UV radiation on the incubator. This also may
cause cracks in the acrylic parts.