Care
127
Operating Instructions Caleo, Software 2.n
Disassembly
1 Lift trough out.
2 Remove fan.
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect all surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
z Turn trough over.
3 Lift air guide plate to the side to disinfect and clean.
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect the surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
Chassis/incubator body
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect the surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
z Remove any dirt near the openings of the sensor unit.
CAUTION !
Do not allow any moisture to enter the sensor unit.
Do not disinfect sensor unit by immersion or spraying.
Sensor damage may result.