Care
126
Operating Instructions Caleo, Software 2.n
Disassembly
Mattress and bed
z Remove mattress from bed.
1 Remove bed.
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect all surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
Remove x-ray drawer
z Using the recessed handle, pull x-ray drawer out as far as it
will go.
2 Tilt drawer upwards and pull it out of the unit.
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect all surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
Remove trough
3 Press both catches inwards and pull the trough upwards.
CAUTION !
Take care not to damage the sensor unit when removing
patient bed.