Care
125
Operating Instructions Caleo, Software 2.n
Disassembly
Open front doors
1 Turn the two locking knobs inwards as far as they will go and
fold down the front door.
z Fold out the hinged double walls to clean them.
z Open the side doors in the same way.
z Remove visible soiling with a disposable cloth and
detergent.
z Wipe-disinfect all surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
U-grommets
z Remove all U-grommets and access port seals.
z Clean parts in a parts washer at 93
o
C (200
o
F)*.
z Push the drainage module** out sideways by the drainage
clip.
z Push the drainage clip off the drainage unit.
z The parts can be disinfected in a parts washer at 93
o
C
(200
o
F)*.
* Observe national and international standards regarding
procedures for cleaning, disinfection, and sterilization.
CAUTION !
U-grommets cannot be autoclaved at 134
o
C (273
o
F).
** Available option