Care
124
Operating Instructions Caleo, Software 2.n
Disassembly
1 Tilt the sleeve to one side and pull it out.
2 Tilt the centring lugs to one side and pull them out.
z Remove obvious soiling with a disposable cloth.
z Wipe-disinfect the double wall surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
z All other parts can be disinfected in a parts washer at 93
o
C
(200
o
F)*.
Remove canopy
3 Grasp handles on sides of canopy with both hands.
4 Keeping canopy level, lift it vertically off its supports.
z Remove obvious soiling with a disposable cloth and
detergent.
z Wipe-disinfect all surfaces.
z After allowing disinfectant to take effect (see manufacturer's
directions for prescribed exposure times), wipe surfaces
with a clean, damp, disposable cloth, then rub dry.
Remove the canopy support
The support (2M 51 141) is only mounted on the pillar
elements (2M 51 154 and 2M 51 156).
5 Move the support to a vertical position and remove it.
z The support can be disinfected in a parts washer at 93
o
C
(200
o
F)*.
* Observe national and international standards regarding
procedures for cleaning, disinfection, and sterilization.