Operating Instructions Caleo, Software 2.n
Before Reusing With a Patient
Check that the side door latches securely
– Perform this test on both side doors
z Open side door slightly.
1 Then, push door back into its closed position. Turn the two
knobs outwards until they tangibly engage in the horizontal
position.
If the side door fails to remain engaged or if the red catch
remains visible:
z Take unit out of service.
Check that the double wall is firmly seated
2 Check that the double wall is securely attached to the hood.
It should not move when slightly pushed.
If the double wall is not securely attached to the hood or the
double wall or parts of the locking mechanism are damaged:
z Do not use the double wall.
z Take unit out of service.
Check that the canopy is securely seated
3 Grasp handle and open canopy.
4 Tilt canopy back (approx. 60
o
).
WARNING !
Always ensure that both knobs of the side doors are
engaged in position in order to avoid any risk of an infant
falling out of an incubator.
The red catch behind each knob must no longer be
visible!