Care
133
Operating Instructions Caleo, Software 2.n
Before Reusing With a Patient
Check all four corners of the canopy for any damage to the
hinge.
z Tilt canopy up slightly, and
1 check canopy prop.
z Lower canopy back.
– Perform check from both sides.
If canopy does not open and close properly:
z Take unit out of service.
2 Grasp handles on the sides of the canopy with both hands.
3 Lift canopy horizontally off the side windows.
If the canopy holders are damaged:
z Take unit out of service.
Check that trough is securely seated
z Remove canopy.
z Remove mattress.
z Remove bed.
4 Check catches for trough.
z Place bed on trough.
z Place mattress on bed.
z Re-install canopy.
If catches for the trough appear damaged:
z Take unit out of service.